Quality Inspection Report of DXWD-60 Electromagnet

Home    News    Quality Inspection Report of DXWD-60 Electromagnet

DXWD-60 laboratory electromagnet  DXWD-60 laboratory electromagnet  DXWD-60 laboratory electromagnet

Overall Dimensions: 426mm x 265mm x 304mm
Coil spacing: 80mm
Pole diameter: 60mm
Pole surface diameter: 30mm
Air gap range: 0-80mm
Total equipment weight: 86kg

 

 

Performance test report  
 Introduction  
1. Test purpose  
  Through the analysis of the test results, the evaluation of the product performance is obtained;
Evaluate whether the test product meets the technical requirements;
Provide reference data for product design and after-sales service, and provide a strong basis for leaving the factory.
 

2. Basic information

 
                Product model: DXWD-60
                Product Name: Electromagnet
                Test time: June 27, 2023
3. Testing environment  
  The detection environment is closer to the actual environmental conditions used by the user, including factors such as lead wire length and ambient temperature. The time is selected in the time interval where the fluctuation of the geomagnetic field is small. Before the test, the instruments required for the test were turned on and warmed up in advance and the measurement environment was recorded, as shown in Table 1.  
Measurement environment  
Environment temperature Product initial temperature Lead length Environmental magnetic field  
26.0℃ 26.0℃ 4.0m 0~30000nT  
Table 1 Measurement environment data  
 Magnetic field parameter test  
    1. Using the device  
  DX-F2031 power supply            
  DX-180 Gauss Meter            
    2. Magnetic field measurement  
  This product uses the following measurement methods:          
  This product uses the following measurement methods:          
Magnetic field strength

Measuring the

air gap

Display

current (A)

Display voltage

Voltage (V)

Magnetic field

strength (T)

10mmGap 1 - 0.304  
10mmGap 2 - 0.614  
10mmGap 3 - 0.923  
10mmGap 4 - 1.226  
10mmGap 5 - 1.504  
10mmGap 10 - 2.056  
20mmGap 1 - 0.152  
20mmGap 2 - 0.302  
20mmGap 3 - 0.451  
20mmGap 4 - 0.601  
20mmGap 5 - 0.748  
20mmGap 10 - 1.288  
Table 2 Magnetic field strength data  
Electrical parameter test  
    1. Using equipment  
   Digital Multimeter            
    2. Test method  
 

It is measured by a digital instrument with a bridge circuit and voltammetry measurement.

See the table below for the measurement results.

 
Resistance R measurement

Environment

temperature

 Product

temperature

Measuring

resistance

Resistance R

@20℃

Resistance R

@85℃

26.0℃ 26.0℃ 7.280Ω 7.084Ω 9.203Ω
Table 3 Resistance R data  
   
Conclusion  
 

The magnetic field and electrical performance of this product meet the

requirements of the corresponding technical parameters, and the final performance

of the product is subject to this document.

 

 

 

Related Products

 

Electromagnet      Helmholtz Coil      Solenoid

Control Magnetic Field Generation System