Electromagnet System: DXWD-130 Electromagnet + DX-150 Gaussmeter + DXMP-20/200 Bipolar Programmable Power Supply + Control Software
Components of the Electromagnet System:

1. DXWD-130 Electromagnet
with D60 mm Pole Head (Natural Cooling Type) and Wheel Base
| Test Report | |||||||
| Introduction | |||||||
| 1)Test aim | |||||||
| Through the analysis of the test results, the evaluation of the product performance is obtained; Evaluate whether the test product meets the technical requirements; Provide reference data for product design and after-sales service, and provide a strong basis for delivery. |
|||||||
| 2)Basci information | |||||||
| Item: | DXG03260321-084A | Model | DXWD-130 | ||||
| Name: | Electromagnet | Test Date: | May 25,2026 | ||||
| 3)Test Condition | |||||||
| The selection of the detection environment is closer to the actual environmental conditions used by the user, including factors such as lead wire length and ambient temperature. The time is selected in the time interval with less fluctuation of the geomagnetic field. Before the test, turn on the instrument required for the test to warm up and record the measurement environment, as shown in Table 1. | |||||||
| Test condition | |||||||
| Temperature of surrounding | Temperature of product | Guide line length | Surrounding Magnetic | ||||
| 21.0℃ | 21.0℃ | / | 0~30000nT | ||||
| Table 1 | |||||||
| Magnetic field parameter | |||||||
| Magnetic field strength | |||||||
| Air gap | Display current (A) | Display voltage(V) | Magnetic field strength(Gs) | ||||
| 35mm | 3 | 28.42 | 3474.4 | ||||
| 6 | 56.84 | 6875.9 | |||||
| 9 | 85.31 | 10107.1 | |||||
| 12 | 113.91 | 12541.1 | |||||
| 15 | 142.75 | 14132.3 | |||||
| Table 2 Magnetic field strength data | |||||||
| Electrical parameters test | |||||||
| 1)Devices used | |||||||
| UT71A Digital Multimeter,TH2829 Impedance Analyzer,DMM6500 Multimeter | |||||||
| 2)Test method | |||||||
| Measured with digital instrument with bridge circuit, voltammetry. The measurement results are shown in the table below. | |||||||
| Resistance R test | |||||||
| Surrounding Temperature | Temperature of product | Test resistance | ResistanceR@20℃ | ResistanceR@85℃ | |||
| 21.0℃ | 15.0℃ | 9.472Ω | 9.429Ω | 12.248Ω | |||
| Table 3 Resistance data | |||||||
| Temperature increasing test | |||||||
| 1) Test method | |||||||
| This product adopts the following measurement methods: | |||||||
| Using magnetic field temperature rise measurement: Given the target magnetic field current required by the user for maximum operating time and temperature | |||||||
| Measure the overall temperature, time, and current data of the coil. The measurement results are shown in the table below | |||||||
| Temperature increasing test | |||||||
| Usage current (A) | Working time (min) | Product temperature (℃) | |||||
| 12 | 0 | 20.2 | |||||
| 5 | 28.8 | ||||||
| 10 | 37.6 | ||||||
| 15 | 45.7 | ||||||
| 20 | 53.5 | ||||||
| 25 | 61.1 | ||||||
| 30 | 68.8 | ||||||
| 35 | 75.8 | ||||||
| 40 | 82.6 | ||||||
| 45 | 89.4 | ||||||
| 50 | 95.8 | ||||||
| 55 | 102.3 | ||||||
| 60 | 108.5 | ||||||
| Note: | The max.continusouly working time of electromagnet 2ith 1.2T@35mm and 12A is 60 minutes | ||||||
| Table 4 Relationship data between temperature, working time, and current | |||||||
| Conclusion | |||||||
| The magnetic field and electrical performance of the product meet the corresponding technical parameters, and the final performance is subject to the document. | |||||||
2. DX-150 Gaussmeter
The Test Report of DX-150 Gaussmeter
Before leaving the factory, all gaussmeter of the company are given the measurement of standard with the nuclear magnetic resonance instrument PT2-26(No.0073) and the standard parallel ring magnetic field generating coil, and the JJF1832-2020 (1Mt-2.5T) Tester meter verification regulations. NIM-ZY-DL CX-209 Constant weak magnetic field calibration inspection as calibration standard.
|
Magnetic field type |
Standard instrument |
China Academy of Metrology |
Test range |
Accuracy |
Remark |
|
Strong magnetic field |
MRI PT2026 |
DCcx2019-00220 |
0.04T-2.0T |
0.0026% |
No |
|
Weak magnetic field |
Standard paralleling ring magnetic field generating coil |
DCcx2020-00366 |
0.001T-0.1T |
0.007% |
Coil constant 10.60839mT/A |
|
Type: |
Tesla/Gauss meter |
||||||||||||||||||||||||||||||||||||||||||||||||
|
Model: |
DX-150 with MCDX800F |
||||||||||||||||||||||||||||||||||||||||||||||||
|
SN: |
DXG03260321-084B |
||||||||||||||||||||||||||||||||||||||||||||||||
|
Date: |
6/1/2026 |
||||||||||||||||||||||||||||||||||||||||||||||||
|
Temperature: |
22℃ |
||||||||||||||||||||||||||||||||||||||||||||||||
|
Humidity: |
45% |
||||||||||||||||||||||||||||||||||||||||||||||||
|
File Number: |
DX-TFCF000850 |
||||||||||||||||||||||||||||||||||||||||||||||||
|
Test Item: Uniform magnetic field standard Test range: 0-100mT Validity: 12 months |
|||||||||||||||||||||||||||||||||||||||||||||||||
|
Test results
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
There will be a certain precision drift after long use. In order to ensure that the reading is within the accuracy range, it is recommended that you send it back to the manufacturer or measurement department for calibration after one year of use. |
|||||||||||||||||||||||||||||||||||||||||||||||||
3. DXMP-20/200 Bipolar Programmable Power Supply
|
TEST REPORT |
|||||
|
Model: DXMP-20/200 |
Serial No.: P202604027 |
Contract No.: PO4500559883 |
|||
|
Parameters |
Output current: ±20 A Output voltage: ±200 V Input voltage: Single Phases 220VAC±10% Cooling form: Forced air cooling ☑ Water cooling □ |
||||
|
Environmental conditions |
Temperature: 23 ℃ Relative Humidity: 36 % |
||||
|
Measurement Instrument |
Keithley DMM7510,DCCT |
||||
|
Test Items |
Inspection Results |
Methods |
Results |
||
|
Appearance |
☑Yes / □No |
Observe whether the appearance of power supply is intact |
☑ PASS □ NG |
||
|
Sign |
Sign (☑Yes / □No) Matched (☑Yes / □No) |
Check whether the serial number of the instrument is matched with the model |
☑ PASS □ NG |
||
|
Panel |
Indicator (☑Yes / □No) Screen display (☑Yes / □No) Touch operation (☑Yes / □No) |
Check whether the following items are normal:Indicator, Screen display, Touch operation |
☑ PASS □ NG |
||
|
Terminals |
☑Yes / □No |
Check whether the terminals have been connected firmly |
☑ PASS □ NG |
||
|
Communication |
RS232 (□Yes / □No / □ N/A) RS485 (□Yes / □No / ☑ N/A) USB (☑Yes / □No / ☑ N/A) LAN (☑Yes / □No / □ N/A) |
Send the "*IDN?" instruction to the instrument, check whether the return data is correct |
☑ PASS □ NG |
||
|
Protection |
Overheat (☑Yes / □No / □ N/A) Overcurrent (☑Yes / □No / □ N/A) Power down (☑Yes / □No / □ N/A) Condensation(□Yes / □No / ☑ N/A) Water-Break (□Yes / □No / ☑ N/A) |
Unplug the temperature switch, Overheat protection; Short-circuit condensation detection, Condensation protection; Close cooling water, Water-Break protection. |
☑ PASS □ NG |
||
|
Output current accuracy |
The current display on the LCD screen: 1. 20.000 2. -20.000 Measured current: 1. 20.0003 2. -19.9998 Accuracy: ±(0.01%+1mA) |
Using the DCCT and Keithley DMM7510 to measure the output current. |
☑ PASS □ NG |
||
|
Output current stability |
Set current: 20A Stability: ±21PPM |
After the power supply is preheated for 30 minutes, the current data shall be collected continuously by software for 1 hour.Then calculate the stability (ppm / h). |
☑ PASS □ NG |
||
|
Traverse zero crossing smoothly |
☑Yes / □No |
Adjust the output current in small steps through the zero crossing, Observe whether there is a sudden change in the current |
☑ PASS □ NG |
||
|
Normal operation |
Constant current test: ☑Yes / □No Scanning current test: ☑Yes / □No |
Make the power supply in working 2 hours at the positive and negative maximum current respectively. Make the power supply in working 4 hours back and forth between positive and negative maximum current by software. |
☑ PASS □ NG |
||
|
Conclusion |
☑ PASS □ NG |
||||
|
Tested By: Robin Date: 2026.5.20 Checked By: Peter Date: 2026.5.21 |
|||||
4. Closed Loop Control Software

Video
Installation and Commissioning of the Electromagnet System