DX-2012MQ Magnetic Moment Testing Equipment: Installation and Testing

Introduction
The connection between magnetometers and Helmholtz coils to test the quality of rare earth materials has become the most popular mutual inspection method among enterprises. Using one-dimensional coils to test axially magnetized cylindrical, circular, and square rare earth permanent magnet samples, and configuring software DX-2012MQ can not only accurately test the total magnetic moment M (more accurately, Mz) of the material, but also for standard shaped NdFeB rare earth permanent magnets, such as square, cylindrical, and circular rings, the BH demagnetization curve under room temperature conditions is close to a straight line, which can accurately calculate the three important parameters representing magnetic performance Br, Hcb, and BHmax, meeting the rapid judgment of material magnetic energy quality.
Features
Designed with a medical-grade 3.5-inch TFT full-color display for more precise 6-digit numerical readouts.
Utilizes an intelligent drift adjustment dynamic algorithm, with the entire drift adjustment process requiring no manual intervention.
One-button measurement of magnetic moment (M), with automatic calculation of values such as Br, Hcb, and BHmax.
Features upper and lower limit sorting functions, displaying OK/NG sorting results.
Equipped with an automated communication interface for fast, automated magnetic flux testing and sorting.
Includes a standard USB interface, with accompanying measuring software for forming an automated measurement system to enable large-scale data storage.
Technical parameters
|
Item |
Specifications |
|
Power Supply |
AC 110V/220V ±10%, 50/60Hz, 1A. According to the site condition |
|
Display |
TFT color LCD, 6-digit display |
|
Dimensions |
307 mm × 260 mm × 120 mm (W × D × H) |
|
Operating Environment |
Temperature: 0°C–50°C; Relative Humidity: 40%–85% RH |
|
Thermal Stabilization Time |
Approximately 20 minutes after connecting the signal input device, until the drift becomes stable |
|
Magnetic Flux Ranges |
10 mWb / 20 mWb / 50 mWb / 100 mWb (four ranges, expandable) |
|
Minimum Resolution |
0.1 μWb (applicable to all ranges) |
|
Zero Drift Adjustment |
Intelligent automatic zero-drift adjustment; no manual operation required |
|
Integral Drift |
≤1 μWb/min or 0.01% RG/min (whichever is greater) |
|
Accuracy |
Br: 0.6%; Hcb: 1.5%; BHmax: 1.5% Accuracy specifications are based on comparative measurement results against the DX-2012H permanent magnet measuring instrument. |
|
Repeatability |
≤0.3% RD + 10 μWb (for ranges ≥50 mWb) |
|
Sorting Function |
Upper and lower sorting limits can be set manually. The instrument automatically displays red/green indicators for NG/OK results. |
|
Communication Interface |
Type-B USB interface |
|
Automation Interface |
Plug-in interface for automated testing (optional) |
|
Digital Output |
RS232C interface (optional); USB interface (for function calls and data retrieval; optional) |
|
Analog Output |
10V corresponding to full-scale range; plug-in interface (optional) |
|
Measurement Software |
Compatible with Windows XP, Windows 7, and Windows 10 (32-bit/64-bit). Measurement data can be saved as text files and viewed using standard office software, including any version of Excel, Word, or Notepad. |
|
Notes: RD = Reading value; RG = Current measurement range value |
|
Standard Configuration:
|
No. |
Item Name |
Model |
Quantity |
|
1 |
Test Main Unit |
DX-2012MQ |
1 |
|
2 |
Accompanying Software |
Accompanying Software |
1 |
|
3 |
One-Dimensional Helmholtz Coil |
According to the requirement |
1 |
|
4 |
Certificate of Conformance & Warranty Card |
Provided by Dexing |
1 |
Note: The above standard configuration is for reference only. The actual product list will be based on the quotation item list.
Optional Accessories or Customization Services:
|
No. |
Item Name |
Model |
|
1 |
One-Dimensional Helmholtz Coil |
DXCT-110D |
|
2 |
One-Dimensional Helmholtz Coil |
DXCT-160D |
|
3 |
One-Dimensional Helmholtz Coil |
DXCT-200D |
|
4 |
One-Dimensional Helmholtz Coil |
DXCT-315D |
|
5 |
One-Dimensional Helmholtz Coil |
DXCT-400D |
|
6 |
Automation Interface Module |
Custom Automation Interface Module (NG\OK Output) |
|
7 |
Digital Output Module |
Custom RS232C Interface Service |
|
8 |
Analog Output Module |
Custom Full-Scale ±10V Analog Output Signal |
Inspection
|
Sample Parameters: |
No: DS10083 Type: cylinder Materials: SmCo Diameter:9.95mm Thickness:9.95mm |
|
Test Environment: |
Room Temperature:25℃ Relative Humidity:46 % |
|
Basis of Measurement Devices |
DX-2012MQ with test coils |
|
Calibration Basis: |
Derive the magnetic test method for permanent magnet (hard magnet) materials according to GB-T 3217 |
|
Results of Calibration |
On page 2 |
Test Condition: DX-2012MQ+DXHC-315
Test Results
|
No. |
Item |
Test Result |
Uncertainty k=2 |
|
1 |
M(uWb*m) |
0.84236 |
1.5% |
|
2 |
Br(T) |
1.10130 |
1.5% |
|
3 |
Hcb(kOe) |
10.4886 |
1.5% |
|
4 |
(BH)max(MGOe) |
28.8781 |
2.5% |
Video